ATC71
Airflow and Temperature Control Plant
The
Mechanical Setup
An inner aluminium tube including an heating element as well as
an axial fan with a regulating flap is mounted on top of an acrylic
base plate.
A differential pressure transmitter, a PT100 temperature sensor with
transmitter as well as an electrical connector box are mounted at the
base plate.
|
|
- Dimension
Sizes and Weight of the System:
- Length
: 545 mm
- Depth
: 260 mm
- Height
: 140 mm
- Weight
: 5,4 kg
- Measuring
Range of the Transmitters:
- Temperature
transmitter : 25 ... 75 °C
- Pressure
transmitter : 0 ... 5 mbar
|
- Technical
Data:
- Max.
heater power : 116 W
- Velocity
flow range : 0 ... ~1,5 m/s
- Tube
length : 340 mm
- Tube
inner diameter : 50 mm
|
Actuator
The actuator box mainly contains a power supply, modules to control
the fan as well as the heating element and measurement output signals
for temperature and airstream. The electronic circuits are housed
in a robust 19" housing.
|
- Dimension
Sizes and Weight of the Actuator:
- Length
: 470 mm
- Depth
: 345 mm
- Height
: 150 mm
- Weight
: 9 kg
- Power
Supply:
- Mains supply : 230 V AC
- Consumption : 150 VA
With the ATC71, either current inputs / outputs
or voltage inputs / outputs can be used.
- Inputs:
- Fan control : 4 ... 20 mA, 0 ... 10V
- Heating control : 4 ... 20 mA, 0 ... 10V
- Input resistance : 200 Ohm
- Overheat
Protection
A bimetallic thermorelay
switches off the heating element at approx. 90 °C. The
heating power is switched off for fan control signal less
than 4,15 mA (2,08 V).
|
- Outputs:
- Measured
signal from temperature sensor : 4 ... 20 mA, 2 ... 10 V
- Measured
signal from pressure sensor : 4 ... 20 mA, 2 ... 10 V
- 2
supply voltages : 24 V DC
- Max.
working resistance : 500Ω
- Binary
signal for flap position : Potential-free contact (Max. supply
24V)
- Binary
signal heat capacity switch : Potential-free contact (Max.
supply 24V)
- Potentiometer:
- internal
fan control : 0 ... 100 %
- internal
heating control : 0 ... 100 %
|
Technical
data are subject to change (Date 8-November-2024) |